Step-One Plus

Location: Main Tzabam, building C, room 004

Booking

Description

Real-Time PCR System that offers Fast runs (<40 min) and standard runs (<2 hr). Innovative VeriFlex™ Block for convenient PCR optimization and includes software capable of collecting melt curve data for High Resolution Melt (HRM) applications.

Features/Specifications

VeriFlex™ Block 25°C (5°C zone-to-zone)
Thermal Cycling System Peltier-based system
Block Format 96-well block
Compatible Consumables • 96-well (0.1 mL) plates with optical flat caps
• 96-well (0.1 mL) plates with optical adhesive cover
• 8-tube (0.1 mL) strips with optical flat caps
• Individual (0.1 mL) tubes with optical flat caps
Supported Volumes 10–30 µl
Note: For standard curve experiments, 40 µl in standard mode is validated.
Sample Ramp Rate • Fast mode: +/- 2.2 °C/sec
• Standard mode: +/- 1.6°C/sec
Peak Block Ramp Rate 4.6 °C/sec
Temperature Range 4°C-100°C
Temperature Accuracy +/-0.25°C (35°C to 95°C) of setpoint/display temperature, measured at 3 minutes after clock start.
Temperature Uniformity +/-0.50°C, 30 seconds after clock start.
Melt Curve Resolution As small as 0.1°C
Optical System Blue LED excitation source, four-emission filters, and photodiode
Calibrated Dyes at Installation FAM™, SYBR® Green I, VIC®, JOE™, NED, TAMRA, ROX™ dyes
Passive Reference Dyes ROX™ or any calibrated dye. Option exists to select no passive reference.
Data Collection Data collected in all 3 filters for all wells regardless of plate setup.
Quantitative PCR Run Time Standard cycling: Less than 2 hrs
Fast cycling: Less than 40 minutes
  • The StepOnePlus™ System supports many Real-Time PCR applications, including the following:
    • SNP Genotyping
    • Gene Expression Analysis
    • MicroRNA Expression
    • Protein Expression
    • Translocation Analysis
    • Gene Detection
    • Viral Load Analysis

The included StepOne™ Software supports a variety of analysis methods, including the following:

  • Standard Curve (absolute quantitation)
  • Relative Standard Curve
  • Comparative Ct (ΔΔCt) (relative quantitation)
  • Genotyping and Presence⁄Absence
  • Melt Curve Analysis (as a standalone application)
  • High Resolution Melting (as a standalone application)

Reservation

BookIt application

Tutoring

First time users must order a tutoring session.

Site

https://products.appliedbiosystems.com/ab/en/US/adirect/ab?cmd=catNavigate2&catID=604109&tab=DetailInfo