Technical details and protocols Optima

Approximate Acceleration Time:

10 acceleration profiles

Approximate Deceleration Time:

11 deceleration profiles

Drive Cooling:

Air-cooled

Electrical Requirements:

200–240V, 30 A, 50/60 Hz

g Force:

694,000 x g

Max Speed:

90,000 rpm

Maximum Heat Dissipation into Room under Steady-State Conditions:

3400 BTU/hr (1.0 kW)

Noise Level 1 Meter in Front of Centrifuge:

<51 dBa

Refrigeration System:

Thermoelectric—no CFCs, ODCs

Set Speed:

1,000 to 90,000 rpm in 100-rpm increments

Set Temperature:

0 to 40° C in 1° C increments

Set Time:

Up to 999 hours, 59 minutes

Speed Control:

±2 rpm of set speed (above 1,000 rpm)

Temperature Control:

±0.5° C of set temperature
±0.8° C when using SW 32 swinging bucket rotor

Ambient Temperature Range:

10° to 35° C