Technical details and protocols Avanti

Features/Specifications

Approximate Acceleration Time:

Two profiles--MAX (full acceleration to set speed) and SLOW

Approximate Deceleration Time:

Three profiles--MAX, SLOW, and OFF (coast)

Electrical Requirements:

50 Hz, 180 to 253 VAC, 24 A; 60 Hz, 187to229/216to264 VAC, 24 A

Friction Reduction System (FRS):

190 mm (7.5 in) Hg

Maximum Heat Dissipation into Room under Steady-State Conditions:

2.0 kW (6,900 Btu/hr)

Noise Level 1 Meter in Front of Centrifuge:

< 64 dBA

Set Speed:

Actual rotor speed ± 50 rpm of set speed

Set Temperature:

-10° C to +40° C in 1° C increments

Set Time:

to 99 hr 59 min or continuous (hold)

Speed Range:

300 to 21,000 rpm

Temperature Control:

± 2° C of set temperature (after equilibration)

Ambient Temperature Range:

15° C to 35° C