Tecan infinite M1000 Pro

Location: Building A,  Robot tzabamon

Booking

Description

Absorption, fluorescence, luminescence, and alpha- screen. The Infinite M1000 PRO offers broad wavelength selection, from UV to NIR, with flexible bandwidth adjustment for optimization of fluorescence intensity based applications.

Features/Specifications

Wavelength selection
  Premium Quad4 Monochromators
  Absorbance Fluorescence
Bandwidth Fixed, 5 nm Adjustable = 300 nm: 2.5 – 10 nm;
> 300 nm: 5 – 20 nm
Wavelength accuracy ± 0.3 nm = 300 nm: ± 0.5 nm;
> 300 nm: ± 1 nm
Wavelength reproducibility ± 0.3 nm = 300 nm: ± 0.5 nm;
> 300 nm: ± 1 nm

 

Wavelength range
  Absorbance Fluorescence FP
  230 – 1,000 nm Ex: 230 – 850 nm; Em: 280 – 850 nm Ex (LED):
470, 530, 590, 635 nm
Em (Mono):
280 – 850 nm

 

Detectors
Absorbance UV Silicon photodiode, 4-channel parallel reading
Fluorescence (incl. FP) Extended wavelength (UV and far red)
Low dark current PMT
Luminescence Photon counting low dark current PMT

 

Plate formats
  6 – 1536-well plates, NanoQuant Plate

 

Temperature control
  Ambient +4 to 42 °C

 

Injection volume Variable; 1 µl increments
Syringe size Selectable: 500, 1000, 2500 µl
Pump speed 100 – 300 µl/sec
Dead volume 100 µl including pump back

 